外骨骼的很不错的国外论文
Sizes of all links, attachments and distances are designed using human anthropometry with 95th percentile. The palm holder and wrist holder are designed to wear easily and both have cushion material on the inner surface to have a comfortable wear.
Usually, the movable range of wrist is 25o-45o in ulnar deviation, 15o-25o in radial deviation, 50o- 70o in extension and 65o-80o in flexion [8]. Also the limitation of the movable range of forearm pronation/supination motion is 50o-80o in pronation and 80o-900 in supination.

(a) Wrist ulnar/radial deviation (b) Wrist flexion/extension

(a) Half -sectional top view

(c) Forearm pronation/supination
Fig. 4 Movable range of the robot.
TABLE 1
Actuation of the three axes.
Axis
Wrist
flexion/extension deviation pronation/supination
(Max.Torque) 0.58(1.38)Nm 0.58(1.38)Nm 4.2 (8.1) Nm
Motor Type
Harmonic drive, super mini RH-5A-4402- 036A0 RH-5A-4402-036A0
RH-8D-6006-E0360D0
(b) Front view
Fig. 3 Forearm cover
Considering the minimally required motion in daily activities and the safety of the user, the limitation of the wrist motion of the exoskeleton is decided to be 30o in ulnar deviation and 20o in radial deviation, 50o in extension and 60o in flexion and that of the forearm motion of exoskeleton is decided to be 60o in pronation and 80o in supination. The movable range of the exoskeleton is shown in Fig. 4. B Sensors and Actuators
All three motors are harmonic drive, super mini RH series DC servo actuators with open collector type incremental encoders. Details of the actuation of the three axes are listed in TABLE.1.


The three axis force sensor is attached between the distal end of the link-3 of the robot and palm attachment of the robot user. Also strain gauges are attached on the connecting beams of link-2 and wrist holder as shown in Fig.3 (b). These force/torque sensors are used to perform force control of the robot.
C Power Transmission
The gear drive systems are used to transmit the power for each joint. A spur gear pair (gear ratio-1:3) is used to transmit the power for the pronation/supination motion. The lager gear is attached to the inner hollow cylinder of the forearm cover by removing the cylindrical piece of material to obtain a hole. Bevel gear pairs (gear ratio-1:2) are used to transmit the power for the wrist ulnar/radial deviation and flexion/extension motions. Modules of each gear pairs are selected to obtain minimum backlash.