平面平晶检测平面度
no attempt should be made to maintain a wedge. A band pattern such as shown in figure 3 will appear.

The bands are to be interpreted as contour lines on a map, the interval being 11.57 millionths of an inch (293.8 nm). Thus the total flatness error is equal to half the band count between points of contact. In the figure shown the 12 bands between high spots indicate a valley 6 bands or 69.42 millionths deep.
Additional Suggestions Regarding Flatness Tests
1. The bands should be viewed as nearly perpendicularly as possible. If viewed from an
increasingly oblique angle the value of 11.57 millionths per dark band can increase
significantly; there will be fewer bands and they tend to straighten.
2. Where an observation from a distance of 10 times the diameter is impractical, a
monochromatic light employing the bent beam principal such as the Van Keuren type L light is recommended.
3. Van Keuren quartz flats distort very little with temperature changes. However, a warm
flat may distort a cold surface. Therefore, all work should be allowed to come to a
common temperature before being tested.
4. The number of bands which appear is not an indication of the flatness of the surface but
relates only to the steepness of the wedge. The bands may be made fewer and father apart by pressing on the optical flat until they are most conveniently spaced for evaluation.
5. A perfectly flat surface will show straight and evenly spaced bands. In this case, a test
with the bands running at right angles to their original position will show curved bands and make it possible to evaluate the curvature.
Effective Band Values
Viewing Band
Angle Value
(micro (degrees) inches)
10 68.6
20 33.8
30 23.1
40 18.00
50 15.1
60 13.4
70 12.3
80 11.8
90 11.6
- 3 -
The determination of the flatness of any ...in the fringes to measure the surface accurately....with repeated use, optical flats can lose their ...
l, the optical path length is change by l and...Typical Measurement Results – Flatness/Shape – ...To measure the temporal coherence, we need detect...
Interferometer is the optical setup which split ...Typical Measurement Results – Flatness/Shape – ...To measure the temporal coherence, we need detect...
amounts, the figure, i.e., flatness of the surface, may be well ...In order to measure surface roughness on polished optical flats, multiple ...
amounts, the figure, i.e., flatness of the surface, may be well ...In order to measure surface roughness on polished optical flats, multiple ...
热门文档