troubleshooting

PNEUMOSLIDE IgM (NSLIDEM)

TROUBLESHOOTING FOR DISTRIBUTORS (provisional)

This document shows guidelines to help distributors to solve and explain possible issues that the end-user might have when performing Pneumoslide IgM.

Possible issues

1. All the wells appear positive in one sample.

Possible causes:

- Inappropriate incubation step. A humid chamber with not enough DI water may produce evaporation in the serum or conjugate which will concentrate increasing the signal. Generally, the desiccation and evaporation is more intense in the periphery of the well. This will make the signal stronger in the central part of the well causing false positives.

- Inappropriate WASH step. This can produce background leading to false positives. - Inappropriate sorbent treatment. This can produce background leading to false positives. For this reason, we recommend a centrifugation step after sorbent incubation to reduce any type of background that can make the visualization of results difficult.

- Inappropriate reading.

- Autoimmunity or cancer disease. When a patient has an autoimmune or cancer disease, antibodies that react with the cells (cellular antibodies or autoantibodies) are produced.

Solutions:

- Inappropriate incubation step (humid chamber with not enough DI water):

Take a chamber with a cover and place folded paper inside, at the bottom (preferable Watman paper, if not cellulose paper).

Use DI water to generously wet the paper to create a humid environment for the incubation. Make sure the bottom surface stays flat.

Place the slides on top of the wet paper and incubate at 37ºC for 90 min. It is extremely important that the incubation lasts for 90 min since it is the time that has been established for the international scientific organizations for the IgMs (that have a large structure) to react with the antigenic epitopes.

- Inappropriate WASH step Immerse the slides in PBS in a wash tray, move manually to rinse the slide and discard the PBS immediately.

Add fresh PBS (never touching the wells directly) and incubate and room temperature with a gentle shaking for 10 min.

Finally, rinse with water to eliminate the salts that can cause background making the observation under the microscope difficult (false positives).

- Inappropriate sorbent treatment (background):