Preparation of single-cell suspensions from mouse heartContents
Preparation of single-cell suspensions from mouse heart
Contents
1. Description
1.1 Background information
1.2 Reagent and instrument requirements
2. Protocol for the dissociation of mouse heart with
Collagenase II treatment
▲ For details on the use of the gentleMACS Dissociator, refer to the gentleMACS Dissociator user manual.
▲ For cell culture experiments subsequent to tissue dissociation, all steps should be performed under sterile conditions.
▲ The weight of one mouse heart amounts to 80–115 mg (female BALB/c mouse, 6–7 weeks old).
▲ One mouse heart or up to four mouse hearts (maximal 400 mg) can be processed per C Tube.1.
Cut mouse heart transversely in two halfs and rinse with cold HBSS.
2. Protocol for the dissociation of mouse heart with
Collagenase II treatment3. References
1. Description
1.1 Background information
Single-cell suspensions are a prerequisite for many experiments, for example to achieve the highest possible purity and recovery during cell separations with MACS® Technology. The gentleMACS Dissociator provides optimized programs to attain single-cell suspensions from various tissues, for example, mouse heart. In combination with C Tubes, the gentleMACS Dissociator allows the automated tissue dissociation in a closed system, enabling sterile sample handling. A single tube or two tubes can be processed in parallel.
This protocol has been developed to obtain endothelial cells and Sca-1+ cells from mouse heart using the gentleMACS Programs m_heart_01 and m_heart_02. Subsets of Sca-1+ cells have been designated as progenitor/stem cells in mouse heart.¹,²1.2 Reagent and instrument requirements
● ● ● ● ●
2. Transfer mouse heart into the gentleMACS C Tube containing
4.7 mL of HBSS. Add 300 µL Collagenase II solution (final Collagenase II concentration: 600 U/mL) and 10 µL DNase I solution (final DNase I concentration: 60 U/mL).
▲ Note: This protocol has been optimized for the use of Collagenase II (Worthington, CLS-2). The use of other enzymes or enzyme concentrations is not recommended.
3. Tightly close C Tube and attach it upside down onto the sleeve of
the gentleMACS Dissociator.
▲ Note: It has to be ensured that the sample material is located in the area of the rotor/stator.
4. Switch on the gentleMACS Dissociator and choose the
gentleMACS Program m_heart_01.5. Run the gentleMACS Program m_heart_01.
6. After termination of the program, detach C Tube from the
gentleMACS Dissociator. 7. Incubate sample for 30 minutes at 37 °C using the MACSmix
Tube Rotator, or turn tube every 5 minutes to resuspend settled tissue fragments.
▲ Note: Operate MACSmix Tube Rotator on permanent run at a speed of approximately 12 rpm.
gentleMACS Dissociator
gentleMACS C Tubes (# 130-093-237)MACSmix Tube Rotator (# 130-090-753)Cell strainer (70 µm mesh size)
(Optional) ART® 1000 REACH pipet tips (Molecular BioProducts, Inc.) for removal of dissociated material from the closed C Tubes.
Red Blood Cell Lysis Solution (10×) (# 130-094-183)
● ●
Hanks’ Balanced Salt Solution with Ca²+ and Mg²+ (HBSS): 1.26 mM CaCl , 0.49 mM MgCl 6H O, 0.41 mM MgSO 7H O, 5.3 mM KCl, 0.44 mM KH PO , 4.17 mM NaHCO , 137.93 mM NaCl, 0.34 mM Na HPO , 5.55 mM Dextrose.9. Keep buffer cold (2 8 °C).
10.
● Collagenase II solution: Prepare a solution with 10,000 U/mL
Collagenase II (Worthington, CLS-2) in HBSS.
11.
● DNase I solution: Prepare a solution containing 30,000 U/mL
DNase I (e.g. DNase I, AppliChem).
12. ● PEB buffer: Prepare a solution containing phosphate-buffered
saline (PBS), pH 7.2, 0.5% bovine serum albumin (BSA), and 2 mM EDTA by diluting MACS BSA Stock Solution (# 130-091- 376) 1:20 with autoMACS Rinsing Solution (# 130-091-222). Keep buffer cold (2 8 °C).
8. Attach C Tube upside down onto the sleeve of the gentleMACS
Dissociator.
▲ Note: It has to be ensured that the sample material is located in the area of the rotor/stator.
Choose and run the gentleMACS Program m_heart_02.After termination of the program, detach C Tube from the gentleMACS Dissociator.
Perform a short centrifugation step to collect the sample material at the tube bottom.
Resuspend sample and apply the cell suspension to a cell strainer (70 µm mesh size) placed on a 50 mL tube.
▲ Note: Dissociated tissue can be removed from the closed C Tube by pipetting through the septum-sealed opening in the center of the cap of the C Tube. Use ART 1000 REACH 1000 μL pipette tips.
140-002-725.02
Miltenyi Biotec GmbH
Friedrich-Ebert-Straße 68, 51429 Bergisch Gladbach, GermanyPhone +49 2204 8306-0, Fax +49 2204 85197macs@miltenyibiotec.de doc.guandang.net
Miltenyi Biotec Inc.
2303 Lindbergh Street, Auburn, CA 95602, USA
Phone 800 FOR MACS, +1 530 888 8871, Fax +1 530 888 8925macs@miltenyibiotec.com
page 1/2
Sample preparation method for isolation of single-cell types from mouse liver...ltered suspensions. 2 Materials and methods 2.3 Isolation of HCs by ...
2.5. Preparation of Lung Single-Cell Suspensions. Lung singlecell suspensions were prepared from the right lung, as detailed previously [15]. Briefly, ...
2.5. Preparation of Lung Single-Cell Suspensions. Lung singlecell suspensions were prepared from the right lung, as detailed previously [15]. Briefly, ...
热门文档