Progesterone Controls
INTENDED USE
The ARCHITECT Progesterone Controls are for the estimation of test precision and the detection of systemic analytical deviations of the ARCHITECT i System (reagents, calibrators and instrument) when used for the quantitative measurement of progesterone in human serum and plasma. Refer to the ARCHITECT Progesterone reagent package insert for additional information.
10

C7K770
STORAGE
ARCHITECT Progesterone Controls are stable until the expiration date
when stored and handled as directed. Do not use past expiration date.
When storing the controls, be certain the vials are securely
sealed. Store at -10°C or colder until first usage. To prepare the controls
for use, thaw completely at room temperature (15-30°C) for 1 to 2 hours, or overnight at 2-8°C. Prior to use, mix THOROUGHLY by low speed vortex or inversion. After each use, immediately return the thawed controls to refrigerated storage (2-8°C). Thawed controls may be stored refrigerated at 2-8°C for up to three weeks. Do not exceed the expiration date printed on the bottle.
[1] US Department of Labor, Occupational Safety and Health
Administration, 29 CFR Part 1910.1030, Bloodborne pathogens.
[2] US Department of Health and Human Services. Biosafety in
Microbiological and Biomedical Laboratories. 5th ed. Washington, DC: US Government Printing Office; January 2007.
[3] World Health Organization. Laboratory Biosafety Manual. 3rd ed.
Geneva: World Health Organization; 2004.
[4] Clinical and Laboratory Standards Institute. Protection of Laboratory
Workers from Occupationally Acquired Infections: Approved Guideline—

Third Edition. CLSI Document M29-A3. Wayne, PA:

Clinical and Laboratory Standards Institute; 2005.ARCHITECT

is a trademark of Abbott Laboratories in various jurisdictions.
Distributed by Abbott LaboratoriesAbbott Park, IL 60064 USAand
ABBOTT 65205 Wiesbaden, Germany
Abbott Ireland
Diagnostics DivisionLisnamuck, LongfordCo. LongfordIreland
+353-43-3331000
CONTENTS


PREPARATION FOR USE

Progesterone The following concentration ranges may be used for individual replicate

control specifications on the ARCHITECT system:


BIBLIOGRAPHY
Each laboratory should establish its own concentration ranges for new
control lots at each control level. This can be accomplished by assaying a minimum of 20 replicates over several (3-5) days. Sources of variation that can

be expected should be included in this study in order to be representative of future system performance. These may include: Multiple stored calibrations Multiple reagent lots Multiple calibrator lots
Multiple processing modules
Data points collected at different times of the day
These results should be applied to your laboratory’s quality control practices.
Diagnostic Use
CAUTION: This product contains human sourced and/or potentially infectious components. Refer to the CONTENTS section of this package insert. No known test method can offer complete assurance that products derived from human sources or inactivated microorganisms will not transmit infection. Therefore, all human sourced materials should be considered potentially infectious. It is recommended that these reagents and human specimens be handled in accordance with the OSHA Standard on Bloodborne Pathogens. [1] Biosafety Level 2[2] or

other appropriate biosafety practices[3],[4] should be used for materials that contain or are suspected of containing infectious agents.
The processed human serum used in the ARCHITECT Progesterone
Controls is nonreactive for HBsAg, HIV-1 RNA or HIV-1 Ag, anti-HCV, disposed of in a safe way.
For information on safe disposal of sodium azide containing materials,
refer to the ARCHITECT System Operations Manual, Section 8.