Innova 2100 system boot up failure 1

g

GEHC China Service

SN00000-X-RAY-TS

SYSTEM: Innova 2100 - Pathfinder

PROBLEM/SYMPTOM

System boot failure, error “process XRDataMgrFront time out after start

component”, “process XRUserIfDLFront time out after start component”

reported on DL monitor during system boot up.

TROUBLESHOOTING HISTORY/HINT

1. Checked RTAC FCIB board status, found yellow LED of config component lit .

2. Exchanged .RTAC FCIB to DL, then found also LED of config component lit, green LED of Fibre Signal and FC Sync do not lit.

3. Replaced RTAC FCIB (PN 5189435) to solve the problem

ROOT CAUSE/SOLUTION

1. System boot up failed due to RTAC FCIB failed.

2. In normal status, green LED of Fibre Signal and FC Sync will light on DL FCIB but do not on RTAC one, we could not judge RTAC FCIB failed by its green LED dark, so if we doubt RTAC FCIB failed, we need to exchanged it to DL location to identify it.

FE name: Jiang Fan Mobile: 18001300800

Page 1 of 1

第1页

你可能喜欢

热门文档

相关文档

  • Innova 3100 Positioner无法启动

    Innova 3100 Positioner无法启动_临床医学_医药卫生_专业资料。g SYSTEM ID: ... Frontal Positioner boot failure( One MCB extended PRD failed) 1050053:Front...

  • Local operating system boot fails when external sto...

    Local operating system boot fails when external storage is attached1_IT/... The failure to bootstrap the OS can be circumvented as follows: From ...

  • 【论文】UP-2410开发板上支持DM9000AE网卡的U-Boot移植

    U-Boot 启动流程分析与S... 32人阅读 2页 1.00UP-2410开发板上支持DM900...

  • 实验一:路由器交换机密码恢复(1)

    boot confreg dir help reset set tftpdnld unset boot up an external process configuration register utility list files in file system monitor builtin command ...

  • Failure record of the meters

    Failure record of the meters_英语学习_外语学习_教育专区。Failure record of...boot the system.察看咪表的时 候,第一次按下唤醒按钮,不能正常启动,多次按...

  • bios1

    Power failure during flash updated Wrong programing while writing to ...boot up ESCD(1) Extended System Configuration Data 记录PNP设备占用的资源 IO...

  • Failure record of the meters

    Failure record of the meters_英语学习_外语学习_教育专区。Failure record of...boot the system.察看咪表的时 候,第一次按下唤醒按钮,不能正常启动,多次按...

站点地图 | 文档上传 | 侵权投诉 | 手机版
本站所有资源均来自互联网,本站只负责收集和整理,均不承担任何法律责任,如有侵权等其它行为请联系我们.
文档下载 Copyright 2013 doc.guandang.net All Rights Reserved.  -  email