Bonaduz, 16 August 2010 th

Reusable flow sensors of Hamilton Medical

Pasteurization for re-processing

Pasteurization is a high level disinfection (HLD) process using hot water at a temperature of 75°C, for a contact time for 30 minutes or longer. High level disinfection process destroys vegetative bacteria, mycobacteria, fungi and eveloped (lipid) and non-eveloped (non-lipid) viruses, but not necessarily bacterial spores. HLD methods, including pasteurization, are indicated to decontaminate / reprocess semi-critical reusable medical devices such as those for anesthesia and respiratory therapy. Due to non-toxicity and moderate heat, pasteurization is particularly suitable for temperature sensitive plastic devices (Key references are provided on the next page).

On this basis, we recommend use of pasteurization, as one applicable method, to re-process the reusable flow sensor (PN 155362) of Hamilton Medical. It is indicated by the internal and external validation that the flow sensor remains functional as specified after 5 reprocessing cycles, at least.

Further, we suggest considering use of advanced washer-disinfectors (WDs) which is based on pasteurization principle. With WDs, cleaning – disinfection – drying process is automatically completed according to the settings (See the next page for the information of WD manufacturers).

After reprocessing, every flow sensor must pass the flow sensor calibration before being used to the next patient. Discard those flow sensors which fail the test twice.

The reprocessing process should be validated microbiologically according to hospitals’ policies of infection control.

CAUTION This recommendation does not apply for any accessory claimed for single use.

With kind regards

Hamilton Medical AG

Dr. Yuan Lei

Frederike Brühschwein Senior Manager Regulatory Affairs fbruehschwein@hamilton-medical.ch Senior Product Manager ylei@hamilton-medical.ch

+41 58 610 13 38

+41 58 610 00 20

Flow-sensor-reprocessing-guide.pdf

Flow-sensor-reprocessing-guide.pdf

Via Crusch 8, 7402 Bonaduz Switzerland Hamilton Medical AG +41 58 610 10 20 info@hamilton-medical.com doc.guandang.net

第1页

你可能喜欢

热门文档

相关文档

  • coldplaner-millingmachin...

    GRADE AND SLOPE CONTROL GRADE SENSOR HEADER CUT The portion of the ...IN-PLACE RECYCLING/ IN SITU REPROCESSING LACING MANUAL DRUM POSITIONING ...

  • 43800-hb-e-aip40

    Any kind of duplication, reprocessing and translation of this document as ...GND Document: 43800-DB-1-4-E-AIP40 17 / 19 GEMAC AIP40 Sensor with ...

  • 瓦里安最新工作站说明书

    325 UV 波长设定完成,其余三项设定包括 Peak sensor、Relay、Miscellaneous 均无...àNext U nR eg is te re 36 d 步骤 18 输入 Reprocessing List 文件名à...

  • 应用过程分析技术(PAT)去优化生物制药生产效率

    Flow Mobile Phase B Innate +/- 2 to 5% variation Flow Sensor Outlet ...(70% recovery per run with 1 reprocessing step of side-cuts results in ...

  • Global percent tree cover at a spatial resolution o...

    10  Page 4 While the MODIS sensor has bands for measuring surface ... and their inclusion in future reprocessing of the land products is ...

  • 生物特征识别及研究现状

    Fea ture e xt ratio n P reprocessing F ace de tect ion Classify desig n Sensor F ace recog nition Fi g. 3 Rid ge endin g a nd rid ge bifur...

  • 声音传感器辐照实验研究_邓骞

    reprocessing system will be supported in result of this research , it is ...sensor in bell incentive 器件所受的总剂量为 在 37 h 辐照实验中, 37 kGy...

  • coldplaner-millingmachin...

    GRADE AND SLOPE CONTROL GRADE SENSOR HEADER CUT The portion of the ...IN-PLACE RECYCLING/ IN SITU REPROCESSING LACING MANUAL DRUM POSITIONING ...

站点地图 | 文档上传 | 侵权投诉 | 手机版
本站所有资源均来自互联网,本站只负责收集和整理,均不承担任何法律责任,如有侵权等其它行为请联系我们.
文档下载 Copyright 2013 doc.guandang.net All Rights Reserved.  -  email