How to derive and use a Color Correction Matrix
to improve display measurements
1. Introduction
The purpose of a Color Correction Matrix (CCM) is to help improve inter-instrument agreement in display and light sources (i.e. emission) measurements. Display measurements are typically done using two instrument types: spectrocolorimeters and colorimeters, the latter usually available at a lower cost. Spectrocolorimeters—such as the Eye-One Pro—measure the display light in separate wavelength bands; these bands are located at fixed intervals, generally five or ten nanometers (nm) apart, between 380 and 730 nm. A spectrocolorimeter can be used to
measure any display type, be it an older Cathode Ray Tube (CRT) display, a standard gamut Liquid Crystal Display (LCD), or a wide-gamut LCD, without any need to correct the measured data according to the display type. Yet, two brands of spectrocolorimeters, or even two different units of the same model, may yield somewhat different responses.
Older colorimeters—such as the DTP94, the Spyder2, the Spyder3, and the Eye-One Display—have just a few sensors whose sensitivities roughly correspond to the Standard Observer XYZ tristimulus responses (a minimum of three sensors are required). These colorimeters were generally optimized for the emission characteristics of CRT phosphors, with some instruments having separate settings for a LCD or a CRT (see for instance the “Display
type” setting in the Display Check tab of the Preferences dialog). Recently introduced colorimeters, such as the i1Display Pro (from X-Rite) and the Spyder4 (from Datacolor), have sensitivities which are closer to the Standard Observer XYZ tristimulus responses. They are also provided with various calibration matrices tweaked for many display types, including wide gamut displays and displays with LED backlights, which are fully supported in PatchTool; these new instruments are designed to be adaptable, in software, to future display technologies. Here again, two brands of colorimeters or two different units of the same model may yield different responses.
With the advent of wide-gamut displays, many have found that their older colorimeters could not accurately
evaluate their display’s performance. A CCM is then a cost-effective solution to extend the operational life of these instruments.
In practice, a Color Correction Matrix can:
Improve the match between measurements taken with two instruments; these instruments can be of the
same model or be different models with different technologies (i.e. spectrocolorimeter vs colorimeter). Correct the response of a colorimeter designed for a specific display type, such as a standard gamut LCD
display with a CCFL backlight (a fluorescent tube), when doing measurements on a display of a different technology (wide gamut, LED backlight, etc.).
Important: Not all colorimeters are inaccurate! In effect, some (very) high end instruments have sensitivities essentially matching the Standard Observer XYZ tristimulus responses and can be used to measure any display
without correction. Unfortunately, the less expensive devices are only accurate for the measurement conditions they were designed for (i.e. to match specific CRT and LCD primaries, or backlights); for other conditions, a CCM can significantly improve their accuracy.